Cited 25 times since 2010 (1.8 per year) source: EuropePMC Journal of electrocardiology, Volume 43, Issue 4, 24 4 2010, Pages 294-301 The spatial QRS-T angle in the Frank vectorcardiogram: accuracy of estimates derived from the 12-lead electrocardiogram. Schreurs CA, Algra AM, Man SC, Cannegieter SC, van der Wall EE, Schalij MJ, Kors JA, Swenne CA

Background and purpose

The spatial QRS-T angle (SA), a predictor of sudden cardiac death, is a vectorcardiographic variable. Gold standard vertorcardiograms (VCGs) are recorded by using the Frank electrode positions. However, with the commonly available 12-lead ECG, VCGs must be synthesized by matrix multiplication (inverse Dower matrix/Kors matrix). Alternatively, Rautaharju proposed a method to calculate SA directly from the 12-lead ECG. Neither spatial angles computed by using the inverse Dower matrix (SA-D) nor by using the Kors matrix (SA-K) or by using Rautaharju's method (SA-R) have been validated with regard to the spatial angles as directly measured in the Frank VCG (SA-F). Our present study aimed to perform this essential validation.

Methods

We analyzed SAs in 1220 simultaneously recorded 12-lead ECGs and VCGs, in all data, in SA-F-based tertiles, and after stratification according to pathology or sex.

Results

Linear regression of SA-K, SA-D, and SA-R on SA-F yielded offsets of 0.01 degree, 20.3 degrees, and 28.3 degrees and slopes of 0.96, 0.86, and 0.79, respectively. The bias of SA-K with respect to SA-F (mean +/- SD, -3.2 degrees +/- 13.9 degrees) was significantly (P < .001) smaller than the bias of both SA-D and SA-R with respect to SA-F (8.0 degrees +/- 18.6 degrees and 9.8 degrees +/- 24.6 degrees, respectively); tertile analysis showed a much more homogeneous behavior of the bias in SA-K than of both the bias in SA-D and in SA-R. In pathologic ECGs, there was no significant bias in SA-K; bias in men and women did not differ.

Conclusion

SA-K resembled SA-F best. In general, when there is no specific reason either to synthesize VCGs with the inverse Dower matrix or to calculate the spatial QRS-T angle with Rautaharju's method, it seems prudent to use the Kors matrix.

J Electrocardiol. 2010 4;43(4):294-301