Cited 30 times since 2005 (1.6 per year) source: EuropePMC Journal of cardiovascular electrophysiology, Volume 16, Issue 10, 1 1 2005, Pages 1097-1103 Validation of ECG indices of ventricular repolarization heterogeneity: a computer simulation study. Van Huysduynen BH, Swenne CA, Draisma HH, Antoni ML, Van De Vooren H, Van Der Wall EE, Schalij MJ

Introduction

Repolarization heterogeneity (RH) is functionally linked to dispersion in refractoriness and to arrhythmogenicity. In the current study, we validate several proposed electrocardiogram (ECG) indices for RH: T-wave amplitude, -area, -complexity, and -symmetry ratio, QT dispersion, and the Tapex-end interval (the latter being an index of transmural dispersion of the repolarization (TDR)).

Methods and results

We used ECGSIM, a mathematical simulation model of ECG genesis in a human thorax, and varied global RH by increasing the standard deviation (SD) of the repolarization instants from 20 (default) to 70 msec in steps of 10 msec. T-wave amplitude, -area, -symmetry, and Tapex-end depended linearly on SD. T-wave amplitude increased from 275 +/- 173 to 881 +/- 456 muV, T-wave area from 34 x 10(3)+/- 21 x 10(3) to 141 x 10(3)+/- 58 x 10(3)muV msec, T-wave symmetry decreased from 1.55 +/- 0.11 to 1.06 +/- 0.23, and Tapex-end increased from 84 +/- 17 to 171 +/- 52 msec. T-wave complexity increased initially but saturated at SD = 50 msec. QT dispersion increased modestly until SD = 40 msec and more rapidly for higher values of SD. TDR increased linearly with SD. Tapex-end increased linearly with TDR, but overestimated it.

Conclusion

T-wave complexity did not discriminate between differences in larger RH values. QT dispersion had low sensitivity in the transitional zone between normal and abnormal RH. In conclusion, T-wave amplitude, -area, -symmetry, and, with some limitations, Tapex-end and T-wave complexity reliably reflect changes in RH.

J Cardiovasc Electrophysiol. 2005 10;16(10):1097-1103