Cited 32 times since 2009 (2.2 per year) source: EuropePMC The Journal of bone and joint surgery. British volume, Volume 91, Issue 9, 1 1 2009, Pages 1134-1141 Ceramic-on-metal bearings in total hip replacement: whole blood metal ion levels and analysis of retrieved components. Isaac GH, Brockett C, Breckon A, van der Jagt D, Williams S, Hardaker C, Fisher J, Schepers A

This study reports on ceramic-on-metal (CoM) bearings in total hip replacement. Whole blood metal ion levels were measured. The median increase in chromium and cobalt at 12 months was 0.08 microg/1 and 0.22 microg/1, respectively, in CoM bearings. Comparable values for metal-on-metal (MoM) were 0.48 microg/1 and 0.32 microg/1. The chromium levels were significantly lower in CoM than in MoM bearings (p = 0.02). The cobalt levels were lower, but the difference was not significant. Examination of two explanted ceramic heads revealed areas of thin metal transfer. CoM bearings (one explanted head and acetabular component, one explanted head and new acetabular component, and three new heads and acetabular components) were tested in a hip joint simulator. The explanted head and acetabular component had higher bedding-in. However, after one million cycles all the wear rates were the same and an order of magnitude less than that reported for MoM bearings. There were four outliers in each clinical group, primarily related to component malposition.

J Bone Joint Surg Br. 2009 9;91(9):1134-1141