Cited 1 times since 2006 (0.1 per year) source: EuropePMC Clinical physiology and functional imaging, Volume 26, Issue 4, 1 1 2006, Pages 220-223 The effect of lung inflation on absolute ventricular volume measurement by conductance. Haney MF, Steendijk P, Johansson G, Biber B

Background

Conductance catheter in vivo ventricular volume measurements during lung ventilation may provide important information on left ventricular (LV) function. Breathing potentially may alter parallel conductance (V(c)), complicating interpretation of the conductance-derived volume signals. The effects of controlled positive pressure lung inflation (PPLI) on measured parallel conductance with dual-field conductance volumetry were investigated.

Methods

Eight anaesthetized pigs were instrumented with an LV dual-field conductance volumetry catheter. V(c) was measured repeatedly, using the hypertonic saline injection method, at three different levels of lung insufflation: continuous positive airway pressure (PPLI) 0, 5, and 10 cm H(2)O.

Results

V(c)s measured at PPLI 0, 5 and 10 cm H(2)O were 70.9 +/- 4.8, 70.7 +/- 5.5 and 70.5 +/- 5.9 ml, respectively. The corresponding uncalibrated end-diastolic volumes (EDV(u)) were 115.5 +/- 7.1, 117.0 +/- 7.5 and 117.5 +/- 7.7 ml, respectively. Mean coefficients of variance for V(c) and EDV(u) at all three PPLI levels were 3.8% and 1.25%, respectively.

Discussion

Several levels of PPLI demonstrated no systematic change in parallel conductance for LV dual-field conductance volume signal. We concluded that lung inflation at these levels does not generate changes in V(c).

Clin Physiol Funct Imaging. 2006 7;26(4):220-223