Cited 6 times since 2020 (1.7 per year) source: EuropePMC JACC. Clinical electrophysiology, Volume 7, Issue 2, 28 4 2020, Pages 197-205 Mini-, Micro-, and Conventional Electrodes: An in Vivo Electrophysiology and Ex Vivo Histology Head-to-Head Comparison. Glashan CA, Beukers HKC, Tofig BJ, Tao Q, Blom S, Mertens B, Kristiansen SB, Zeppenfeld K

Objectives

This study sought to assess the relative effect of catheter, tissue, and catheter-tissue parameters, on the ability to determine the amount of viable myocardium in vivo.

Background

Although multiple variables impact bipolar voltages (BVs), electrode size, interelectrode spacing, and directional dependency are of particular interest with the development of catheters incorporating mini and microelectrodes.

Methods

Nine swine with early reperfusion myocardial infarctions were mapped using the QDot catheter and then remapped using a Pentaray catheter. All QDot points were matched with Pentaray points within 5 mm. The swine were sacrificed, and mapping data projected onto the heart. Transmural biopsies corresponding to mapping points were obtained, allowing a comparison of electrograms recorded by mini, micro-, and conventional electrodes with histology.

Results

The conventional BV of 2,322 QDot points was 1.9 ± 1.3 mV. The largest of the 3 microelectrode BVs (BVµMax) average 4.8 ± 3.1 mV. The difference between the largest (BVμMax) and smallest (BVμMin) at a given location was 53.7 ± 18.1%. The relationships between both BVμMax and BVμMin and between the conventional BV and BVμMax were positively related but with a significant spread in data, which was more pronounced for the latter. Pentaray points positively related to the BVμMax with poor fit. On histology, increasing viable myocardium increased voltage, but both the slope coefficient and fit were best for BVμMax.

Conclusions

Using histology, we could demonstrate that BVμMax is superior to identify viable myocardium compared with BVC and BV using the Pentaray catheter. The ability to simultaneously record 3 BVμs with different orientations, for the same beat, with controllable contact and selecting BVμMax for local BV may partially compensate for wave front direction.

JACC Clin Electrophysiol. 2020 10;7(2):197-205