Cited 73 times since 2016 (9.8 per year) source: EuropePMC Journal of the American College of Cardiology, Volume 68, Issue 19, 1 1 2016, Pages 2047-2055 Long-Term Clinical Outcomes of Subcutaneous Versus Transvenous Implantable Defibrillator Therapy. Brouwer TF, Yilmaz D, Lindeboom R, Buiten MS, Olde Nordkamp LR, Schalij MJ, Wilde AA, van Erven L, Knops RE

Background

Transvenous implantable cardioverter-defibrillators (TV-ICDs) improve survival in patients at risk for sudden cardiac death, but complications remain an important drawback. The subcutaneous ICD (S-ICD) was developed to overcome lead-related complications. Comparison of clinical outcomes of both device types in previous studies was hampered by dissimilar patient characteristics.

Objectives

This retrospective study compares long-term clinical outcomes of S-ICD and TV-ICD therapy in a propensity-matched cohort.

Methods

The authors analyzed 1,160 patients who underwent S-ICD or TV-ICD implantation in 2 high-volume hospitals in the Netherlands. Propensity matching for 16 baseline characteristics, including diagnosis, yielded 140 matched pairs. Clinical outcomes were device-related complications requiring surgical intervention, appropriate and inappropriate ICD therapy, and were reported as 5-year Kaplan-Meier rate estimates.

Results

All 16 baseline characteristics were balanced in the matched cohort of 140 patients with S-ICDs and 140 patients with TV-ICDs (median age 41 years [interquartile range: 30 to 52 years] and 40% women). The complication rate was 13.7% in the S-ICD group versus 18.0% in the TV-ICD group (p = 0.80). The infection rate was 4.1% versus 3.6% in the TV-ICD groups (p = 0.36). Lead complications were lower in the S-ICD arm compared with the TV-ICD arm, 0.8% versus 11.5%, respectively (p = 0.03). S-ICD patients had more nonlead-related complications than TV-ICD patients, 9.9% versus 2.2%, respectively (p = 0.047). Appropriate ICD intervention (antitachycardia pacing and shocks) occurred more often in the TV-ICD group (hazard ratio [HR]: 2.42; p = 0.01). The incidence of appropriate (TV-ICD HR: 1.46; p = 0.36) and inappropriate shocks (TV-ICD HR: 0.85; p = 0.64) was similar.

Conclusions

The complication rate in patients implanted with an S-ICD or TV-ICD was similar, but their nature differed. The S-ICD reduced lead-related complications significantly, at the cost of nonlead-related complications. Rates of appropriate and inappropriate shocks were similar between the 2 groups.

J Am Coll Cardiol. 2016 11;68(19):2047-2055