Cited 9 times since 2012 (0.8 per year) source: EuropePMC Journal of magnetic resonance imaging : JMRI, Volume 36, Issue 6, 22 4 2012, Pages 1470-1476 Evaluation of sampling density on the accuracy of aortic pulse wave velocity from velocity-encoded MRI in patients with Marfan syndrome. Kröner ES, van der Geest RJ, Scholte AJ, Kroft LJ, van den Boogaard PJ, Hendriksen D, Lamb HJ, Siebelink HM, Mulder BJ, Groenink M, Radonic T, Hilhorst-Hofstee Y, Bax JJ, van der Wall EE, de Roos A, Reiber JH, Westenberg JJ

Purpose

To evaluate the effect of spatial (ie, number of sampling locations along the aorta) and temporal sampling density on aortic pulse wave velocity (PWV) assessment from velocity-encoded MRI in patients with Marfan syndrome (MFS).

Materials and methods

Twenty-three MFS patients (12 men, mean age 36 ± 14 years) were included. Three PWV-methods were evaluated: 1) reference PWV(i.p.) from in-plane velocity-encoded MRI with dense temporal and spatial sampling; 2) conventional PWV(t.p.) from through-plane velocity-encoded MRI with dense temporal but sparse spatial sampling at three aortic locations; 3) EPI-accelerated PWV(t.p.) with sparse temporal but improved spatial sampling at five aortic locations with acceleration by echo-planar imaging (EPI).

Results

Despite inferior temporal resolution, EPI-accelerated PWV(t.p.) showed stronger correlation (r = 0.92 vs. r = 0.65, P = 0.03) with reference PWV(i.p.) in the total aorta, with less error (8% vs. 16%) and variation (11% vs. 27%) as compared to conventional PWV(t.p.) . In the aortic arch, correlation was comparable for both EPI-accelerated and conventional PWV(t.p.) with reference PWV(i.p.) (r = 0.66 vs. r = 0.67, P = 0.46), albeit 92% scan-time reduction by EPI-acceleration.

Conclusion

Improving spatial sampling density by adding two acquisition planes along the aorta results in more accurate PWV assessment, even when temporal resolution decreases. For regional PWV assessment in the aortic arch, EPI-accelerated and conventional PWV assessment are comparably accurate. Scan-time reduction makes EPI-accelerated PWV assessment the preferred method of choice.

J Magn Reson Imaging. 2012 6;36(6):1470-1476